Nitrite Determination Assay Kit
-
- Nachweismethode
- Colorimetric
- Detektionsbereich
- 1-100 μM
- Untere Nachweisgrenze
- 1 μM
- Applikation
- Biochemical Assay (BCA)
- Verwendungszweck
- The kit is designed for nitrite determination in biological samples.
- Proben
- Plasma
- Spezifität
- Specific for nitrite determination in plasma samples.
- Produktmerkmale
- The overproduction of nitric oxide may lead to oxidative and nitrosative stress. It has been demonstrated that they enhance the development of a variety of diseases, as well as the ageing process1. Regarding nitrosative stress, high levels of iNOS have be
- Bestandteile
- Plate and the reagents neccesary to perform the assay.
- Benötigtes Material
- Pipettes, reaction tubes, plate reader.
-
-
- Probenmenge
- 50 μL
- Testdauer
- 1 h
- Plattentyp
- Uncoated
- Aufbereitung der Reagenzien
-
No reagent preparation is needed.
- Aufbereitung der Proben
-
Plasma samples may be deproteinized before performing the assay.
- Testdurchführung
-
Standard Preparation: Prepare the calibrate in 1 mL tubes following the Table 1 ( see in booklet) . Use ultrapure water as diluent. Performing the assay: 1. Add 50 μL of samples or standards in each well (96-well plate). 2.Add 50 μL of Reagent A in each well. Incubate for 10 minutes protected from light. 3.Add 50 μL of Reagent B in each well. Incubate for 10 minutes protected from light. 4.Read the absorbance at 540 nm within 30 minutes.
- Ergebnisberechnung
-
Determine average absorbance value of each experimental sample. Determine its concentration by comparison to the Nitrite Standard reference curve (Figure 2 see images). Nitrite (μM) = (?A540nm- intercept)/ slope
- Beschränkungen
- Nur für Forschungszwecke einsetzbar
-
- Konservierungsmittel
- Sodium azide
- Vorsichtsmaßnahmen
- This product contains Sodium azide: a POISONOUS AND HAZARDOUS SUBSTANCE which should be handled by trained staff only.
- Lagerung
- 4 °C
-