Ferritin ELISA Kit

FE Reaktivität: Human Colorimetric Sandwich ELISA 0.6-200 ng/mL Cell Culture Supernatant, Plasma, Serum
Produktnummer ABIN624974
  • Target Alle Ferritin (FE) ELISA Kits anzeigen
    Ferritin (FE)
    Reaktivität
    • 5
    • 4
    • 3
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Human
    Nachweismethode
    Colorimetric
    Methodentyp
    Sandwich ELISA
    Detektionsbereich
    0.6-200 ng/mL
    Untere Nachweisgrenze
    0.6 ng/mL
    Applikation
    ELISA
    Verwendungszweck
    Human Ferritin ELISA Kit for cell culture supernatants, plasma, and serum samples.
    Proben
    Plasma, Cell Culture Supernatant, Serum
    Analytische Methode
    Quantitative
    Spezifität
    This Ferritin ELISA kit shows no cross-reactivity with the following cytokines tested: human Angiogenin, BDNF, BLC, CNTF, ENA- 78, FGF-4, IL-1 alpha, IL-1 beta, IL-2, IL-3, IL-4, IL-5, IL-6, IL-7, IL-8, IL-9, IL-11, IL-12 p70, IL-12 p40, IL-13, IL-15, I-309, IP-10, FGF-4, FGF-6, FGF- 7, G-CSF, GDNF, GM-CSF, IFN-gamma, IGFBP-2, IGFBP-3, IGFBP-4, Leptin (OB), MCP-1, MCP-2, MCP-3, MDC, MIF, MIG, MIP-1 alpha, MIP-1 beta, MIP-1 delta, PARC, PDGF, RANTES, SCF, SDF-1 alpha, TARC, TGF-beta, TIMP-1, TIMP-2, TNF-alpha, TNF-beta, TPO, VEGF.
    Sensitivität
    600 pg/mL
    Produktmerkmale
    • Strip plates and additional reagents allow for use in multiple experiments
    • Quantitative protein detection
    • Establishes normal range
    • The best products for confirmation of antibody array data
    Bestandteile
    • Pre-Coated 96-well Strip Microplate
    • Wash Buffer
    • Stop Solution
    • Assay Diluent(s)
    • Lyophilized Standard
    • Biotinylated Detection Antibody
    • Streptavidin-Conjugated HRP
    • TMB One-Step Substrate
    Benötigtes Material
    • Distilled or deionized water
    • Precision pipettes to deliver 2 μL to 1 μL volumes
    • Adjustable 1-25 μL pipettes for reagent preparation
    • 100 μL and 1 liter graduated cylinders
    • Tubes to prepare standard and sample dilutions
    • Absorbent paper
    • Microplate reader capable of measuring absorbance at 450nm
    • Log-log graph paper or computer and software for ELISA data analysis
    Featured
    Zu unserem meistverkauften FE ELISA Kit
    Top Product
    Discover our top product FE ELISA Kit
  • Applikationshinweise
    Recommended Dilution for serum and plasma samples2 - 10 fold
    Probenmenge
    100 μL
    Plattentyp
    Pre-coated
    Protokoll
    1. Prepare all reagents, samples and standards as instructed in the manual.
    2. Add 100 μL of standard or sample to each well.
    3. Incubate 2.5 h at RT or O/N at 4 °C.
    4. Add 100 μL of prepared biotin antibody to each well.
    5. Incubate 1 h at RT.
    6. Add 100 μL of prepared Streptavidin solution to each well.
    7. Incubate 45 min at RT.
    8. Add 100 μL of TMB One-Step Substrate Reagent to each well.
    9. Incubate 30 min at RT.
    10. Add 50 μL of Stop Solution to each well.
    11. Read at 450 nm immediately.
    Aufbereitung der Reagenzien
    1. Bring all reagents and samples to room temperature (18 - 25 °C) before use.
      2. Sample dilution: If your samples need to be diluted, Assay Diluent A (Item D) should be used for dilution of serum/plasma samples. 1x Assay Diluent B (Item E) should be used for dilution of cell culture supernantants and urine. Suggested dilution for normal serum/plasma: 2-10 fold. Please note that levels of the target protein may vary between different specimens. Optimal dilution factors for each sample must be determined by the investigator.
      3. Assay Diluent B should be diluted 5-fold with deionized or distilled water before use.
      4. Preparation of standard: Briefly spin the vial of Item C and then add 400 µL Assay Diluent A (for serum/plasma samples) or 1x Assay Diluent B (for cell culture supernates/urine) into Item C vial to prepare a 200 ng/mL standard. Dissolve the powder thoroughly by a gentle mix. Pipette 270 µL Assay Diluent A or 1x Assay Diluent B into each tube. Use the 200 ng/mL standard solution to produce a dilution series . Mix each tube thoroughly before the next transfer. Assay Diluent A or 1x Assay Diluent B serves as the zero standard (0 ng/mL). Standard, Item C + 400 µL 180 µL 180 µL 180 µL 180 µL 180 µL 180myl 200 80 32 12 5.12 2.04 0.81 0 ng/mL ng/mL ng/mL ng/mL ng/mL ng/mL ng/mL ng/mL
      5. If the Wash Concentrate (20x) (Item B) contains visible crystals, warm to room temperature and mix gently until dissolved. Dilute 20 ml of Wash Buffer Concentrate into deionized or distilled water to yield 400 ml of 1x Wash Buffer.
      6. Briefly spin the Detection Antibody vial (Item F) before use. Add 100 µL of 1x Assay Diluent B into the vial to prepare a detection antibody concentrate. Pipette up and down to mix gently (the concentrate can be stored at 4 °C for 5 days). The detection antibody concentrate should be diluted 80-fold with 1x Assay Diluent B and used in step 4 of Part VI Assay Procedure.
      7. Briefly spin the HRP-Streptavidin concentrate vial (Item G) and pipette up and down to mix gently before use. HRP-Streptavidin concentrate should be diluted 400-fold with 1x Assay Diluent B. For example: Briefly spin the vial (Item G) and pipette up and down to mix gently . Add 25 µL of HRP-Streptavidin concentrate into a tube with 10 ml 1x Assay Diluent B to prepare a final 400 fold diluted HRP- Streptavidin solution (don't store the diluted solution for next day use). Mix well.
    Testdurchführung
    1. Bring all reagents and samples to room temperature (18 - 25 °C) before use. It is recommended that all standards and samples be run at least in duplicate.
      2. Add 100 µL of each standard (see Reagent Preparation step 2) and sample into appropriate wells. Cover well and incubate for 2.5 hours at room temperature or over night at 4 °C with gentle shaking.
      3. Discard the solution and wash 4 times with 1x Wash Solution. Wash by filling each well with Wash Buffer (300 myl) using a multi-channel Pipette or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot it against clean paper towels.
      4. Add 100 µL of 1x prepared biotinylated antibody (Reagent Preparation step 6) to each well. Incubate for 1 hour at room temperature with gentle shaking.
      5. Discard the solution. Repeat the wash as in step
      6. Add 100 µL of prepared Streptavidin solution (see Reagent Preparation step 7) to each well. Incubate for 45 minutes at room temperature with gentle shaking.
      7. Discard the solution. Repeat the wash as in step
      8. Add 100 µL of TMB One-Step Substrate Reagent (Item H) to each well. Incubate for 30 minutes at room temperature in the dark with gentle shaking.
      9. Add 50 µL of Stop Solution (Item I) to each well. Read at 450 nm immediately.
    Ergebnisberechnung

    Calculate the mean absorbance for each set of duplicate standards, controls and samples, and subtract the average zero standard optical density. Plot the standard curve on log-log graph paper or using Sigma plot software, with standard concentration on the x-axis and absorbance on the y-axis. Draw the best-fit straight line through the standard points.
    Typical Data: These standard curves are for demonstration only. A standard curve must be run with each assay. Assay Diluent A Human Ferritin concentration (pg/mL) O D =4 50 n m 0.01 0.1 1 10 0 0.1 1 10 100 Assay Diluent B Human Ferritin concentration (pg/mL) O D =4 50 n m 0.01 0.1 1 10 0 0.1 1 10 100
    Sensitivity: The minimum detectable dose of Ferritin is typically less than 0.6 ng/mL.
    Recovery: Recovery was determined by spiking various levels human Ferritin into human serum, plasma and cell culture media. Mean recoveries are as follows: Sample Type Average % Recovery Range ( %) Serum 97.32 86-105 Plasma 103.1 93-112 Cell culture media 96.46 85-93
    Linearity: Sample Type Serum Plasma Cell Culture Media 1:2 Average % of Expected 97.78 102.8 105.4 Range ( %) 86-106 91-110 96-113 1:4 Average % of Expected 101.1 75.54 86.67 Range ( %) 91-110 67-81 75-94
    Reproducibility: Intra-Assay: CV<10 % Inter-Assay: CV<12 %

    Testpräzision
    Intra-Assay: CV< 10 % Inter-Assay: CV< 12 %
    Beschränkungen
    Nur für Forschungszwecke einsetzbar
  • Handhabung
    Avoid repeated freeze-thaw cycles.
    Lagerung
    -20 °C
    Informationen zur Lagerung
    The entire kit may be stored at -20°C for up to 1 year from the date of shipment. Avoid repeated freeze-thaw cycles. The kit may be stored at 4°C for up to 6 months. For extended storage, it is recommended to store at -80°C.
    Haltbarkeit
    6 months
  • Vaz-Tostes, Verediano, de Mejia, Brunoro Costa: "Evaluation of iron and zinc bioavailability of beans targeted for biofortification using in vitro and in vivo models and their effect on the nutritional status of preschool children." in: Journal of the science of food and agriculture, Vol. 96, Issue 4, pp. 1326-32, (2016) (PubMed).

    Eghbali, Taherahmadi, Bagheri, Nikanjam, Ebrahimi: "Association between serum ferritin level and diastolic cardiac function in patients with major β-thalassemia." in: Iranian journal of pediatric hematology and oncology, Vol. 5, Issue 2, pp. 83-8, (2015) (PubMed).

    Vanarsa, Ye, Han, Xie, Mohan, Wu: "Inflammation associated anemia and ferritin as disease markers in SLE." in: Arthritis research & therapy, Vol. 14, Issue 4, pp. R182, (2014) (PubMed).

    Grunert, Reutgen, Iwainsky: "[On qualitative demonstration of tuberculostatics in the control of taking of drugs]." in: Zeitschrift für Tuberkulose und Erkrankungen der Thoraxorgane, Vol. 124, Issue 3, pp. 141-8 (PubMed).

  • Target Alle Ferritin (FE) ELISA Kits anzeigen
    Ferritin (FE)
    Andere Bezeichnung
    Ferritin (FE Produkte)
    Synonyme
    1HCH ELISA Kit, BEST:LD36673 ELISA Kit, CG2216 ELISA Kit, Dmel\\CG2216 ELISA Kit, FER1 ELISA Kit, FER1HCH ELISA Kit, Fer-L ELISA Kit, Fer1 ELISA Kit, Fer1HC ELISA Kit, FerHCH ELISA Kit, HCH ELISA Kit, T21 ELISA Kit, T9 ELISA Kit, fer1 ELISA Kit, fer1hch ELISA Kit, l(3)00451 ELISA Kit, l(3)j10B4 ELISA Kit, SFerH-1 ELISA Kit, SOF-35 ELISA Kit, SOF-5L ELISA Kit, NV13011 ELISA Kit, ACP1 ELISA Kit, F8A5.24 ELISA Kit, F8A5_24 ELISA Kit, Ferritin 1 heavy chain homologue ELISA Kit, ferritin light chain ELISA Kit, ferritin ELISA Kit, chromosome 26 open reading frame, human C10orf131 ELISA Kit, hyaluronan and proteoglycan link protein 2 ELISA Kit, NAD(P)-linked oxidoreductase superfamily protein ELISA Kit, Fer1HCH ELISA Kit, H-1 ELISA Kit, YPDSF_1340 ELISA Kit, ftnA ELISA Kit, Ft ELISA Kit, C26H10orf131 ELISA Kit, HAPLN2 ELISA Kit, AT1G60730 ELISA Kit
    Hintergrund
    The Human Ferritin ELISA (Enzyme-Linked Immunosorbent Assay) kit is an in vitro enzyme-linked immunosorbent assay for the quantitative measurement of human Ferritin in serum, plasma, cell culture supernatants and urine. This assay employs an antibody specific for human Ferritin coated on a 96-well plate. Standards and samples are pipetted into the wells and Ferritin present in a sample is bound to the wells by the immobilized antibody. The wells are washed and biotinylated anti-human Ferritin antibody is added. After washing away unbound biotinylated antibody, HRP-conjugated streptavidin is pipetted to the wells. The wells are again washed, a TMB substrate solution is added to the wells and color develops in proportion to the amount of Ferritin bound. The Stop Solution changes the color from blue to yellow, and the intensity of the color is measured at 450 nm. Reproducibility: Intra-Assay: CV<10% Inter-Assay: CV<12%.
    Gen-ID
    2495
    UniProt
    P02794
    Pathways
    Transition Metal Ion Homeostasis