Hemopexin CLIA Kit
-
- Target Alle Hemopexin (HPX) CLIA Kits anzeigen
- Hemopexin (HPX)
-
Reaktivität
- Human
- Nachweismethode
- Chemiluminescent
- Methodentyp
- Sandwich ELISA
- Detektionsbereich
- 0.82 ng/mL - 600 ng/mL
- Untere Nachweisgrenze
- 0.82 ng/mL
- Applikation
- ELISA
- Proben
- Plasma, Serum
- Analytische Methode
- Quantitative
- Spezifität
- This assay has high sensitivity and excellent specificity for detection of Hemopexin (HPX). No significant cross-reactivity or interference between Hemopexin (HPX) and analogues was observed.
- Sensitivität
- 0.39 ng/mL
- Featured
- Zu unserem meistverkauften HPX CLIA Kit
- Top Product
- Discover our top product HPX ELISA Kit
-
-
- Kommentare
-
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
- Testdauer
- 2 - 3 h
- Plattentyp
- Pre-coated
- Protokoll
- The microplate provided in this kit has been pre-coated with an antibody specific to Hemopexin (HPX). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Hemopexin (HPX). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Hemopexin (HPX) level in the sample or standard.,
- Testpräzision
-
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Hemopexin (HPX) were tested 20 times on one plate, respectively
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Hemopexin (HPX) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12% - Beschränkungen
- Nur für Forschungszwecke einsetzbar
-
- Handhabung
- Do not allow to contact skin or eyes. Calibrators, controls and specimen samples should be assayed in duplicate. Once the procedure has been started, all steps should be completed without interruption.
- Lagerung
- 4 °C,-20 °C
- Informationen zur Lagerung
- -20°C. Bring all reagents to room temperature before beginning test. The kit may be stored at 4°C for immediate use within two days upon arrival. Reseal any unused strips with desiccant pack. Minimize freeze/thaw cycles.
- Haltbarkeit
- 4-8 months
-
- Target Alle Hemopexin (HPX) CLIA Kits anzeigen
- Hemopexin (HPX)
- Andere Bezeichnung
- Hemopexin (HPX Produkte)
- Synonyme
- HPX CLIA Kit, HX CLIA Kit, Hpxn CLIA Kit, hx CLIA Kit, wu:fi17f11 CLIA Kit, wu:fi19a10 CLIA Kit, hemopexin CLIA Kit, HPX CLIA Kit, Mmar10_1124 CLIA Kit, Hpx CLIA Kit, hpx CLIA Kit
- Hintergrund
-
Gene Name: Hemopexin
Gene Aliases: Haemopexin, Beta-1B-glycoprotein
- Gen-ID
- 3263
- UniProt
- P02790
- Pathways
- Transition Metal Ion Homeostasis, Regulation of Leukocyte Mediated Immunity, Positive Regulation of Immune Effector Process, Production of Molecular Mediator of Immune Response
-