Dieses Colorimetric ELISA-Kit wurde entwickelt für die quantitative Messung von Human, Maus und Ratte Trastuzumab.
Produktnummer ABIN4886397

Kurzübersicht für Trastuzumab ELISA Kit (ABIN4886397)

Target

Alle Trastuzumab Produkte
Trastuzumab

Reaktivität

Human, Maus, Ratte

Nachweismethode

Colorimetric

Methodentyp

Sandwich ELISA

Detektionsbereich

1.25-40 ng/mL

Applikation

ELISA

Proben

Plasma, Serum
  • Untere Nachweisgrenze

    1.25 ng/mL

    Verwendungszweck

    Quantification of Trastuzumab in biological matrices

    Analytische Methode

    Quantitative

    Spezifität

    Trastuzumab (Herceptin)

    Kreuzreaktivität (Details)

    hIgG1, Rituximab, and Infliximab prepared at 250 ng/mL were assayed and exhibited no cross-reactivity or interference.

    Bestandteile

    Coated microtiter plate, 96 wells
    Calibrator diluent. - 1.8ml
    Calibrator 12ul
    10X wash buffer - 25ml
    Assay buffer - 50ml
    1000X detection reagent - 17ul
    TMB - 12ml
    TMB stop solution - 12ml
    Plate sealers - 3

    Benötigtes Material

    Precision pipettes calibrated to deliver 5-1000μL
    Multi-channel pipette calibrated to deliver 50-200μL
    Plate shaker
    Disposable tips
    Vortex-Mixer
    Distilled or de-ionized water
    Microplate reader capable of reading 450nm with background subtrac
  • Applikationshinweise

    Optimal working dilution should be determined by the investigator.

    Probenmenge

    15 μL

    Testdauer

    2.5 h

    Plattentyp

    Pre-coated

    Protokoll

    The Trastuzumab ELISA kit is designed to measure free Trastuzumab with high specificity and sensitivity . This assay employs the sandwich enzyme immunoassay technique. A precoated anti-Trastuzumab 96 well plate is provided. Calibrator, quality control samples and test samples are pipetted into the appropriate wells. Trastuzumab present in biological matrices is bound by the immobilized capture antibody. After washing away any unbound substances, enzyme linked detection antibody is added to the wells. The plate is washed to remove any unbound antibody-enzyme reagent and a substrate solution is added to the wells for color development. The color development is proportional to the amount of Trastuzumab present in test samples and the concentration is calculated from the standard series.

    Aufbereitung der Reagenzien

    Prepare only the appropriate amount of required reagent on the day of use. Store all reagents as per instructions stated on the label. 1. Wash Buffer (1X) Preparation: Dilute wash buffer concentrate with ultra-pure water 1/10 before use (for example add 10 mL concentrate to 90 mL ultra-pure water). Mix well. 2. Detection Reagent (1X) Preparation: Dilute detection reagent with assay buffer 1/1000 before use (for example add 10 μL concentrate to 10 mL of assay buffer). Mix welll. 3. Preparation of Calibrators: Prepare calibrators with concentrations ranging from 1000 ng/mL to 62.5 ng/mL. The following is an example calibrator curve.

    Probennahme

    This kit is compatible with EDTA-plasma, heparinplasma and serum samples. Samples can be stored at or below -20 °C for up to 1 year.

    Aufbereitung der Proben

    Dilute calibrators and test samples 1/50 with assay buffer (for example add 5μL of prepared calibrator or sample to 245μL of assay buffer). Mix well. Do not store diluted samples.

    Testdurchführung

    This assay employs the sandwich enzyme immunoassay technique. Anti- Trastuzumab is coated onto a 96 well microplate. Calibrator, quality control samples (if desired) and test samples are pipetted into the appropriate wells. Trastuzumab present in biological matrices is bound by the immobilized anti- Trastuzumab antibody. After washing away any unbound substances, enzyme linked antiTrastuzumab antibody is added to the wells. The plate is washed to remove any unbound antibody-enzyme reagent and a substrate solution is added to the wells for color development. The color development is proportional to the amount of Trastuzumab present in test samples. The color development is stopped and the intensity of the color is measured.

    Ergebnisberechnung

    1. Construct a standard curve by plotting the absorbance obtained from each standard against concentration. Use a 4 or 5 parameter curve fit. Alternatively a log-log curve fit may be used. 2. The concentration of the unknowns can be read directly from this standard curve using the absorbance value for each sample. 3. Any sample undiluted or diluted still reading greater than the highest standard should be diluted appropriately with calibrator diluent and retested. If the samples have been diluted, the concentration determined from the standard curve must be multiplied by the dilution factor.

    Testpräzision

    Precision: The precision was determined by analyzing samples prepared at 1000 ng/mL in 6 replicates on 6 different occasions. Intra-assay coefficient of variation (CV) < 10%. Inter-assay CV < 10%.

    Recovery: 1000ng/mL of Trastuzumab was spiked in 10 lots of human serum. Recovery ranges are from 91-117% with an average recovery of 106%.

    Beschränkungen

    Nur für Forschungszwecke einsetzbar
  • Konservierungsmittel

    Without preservative

    Vorsichtsmaßnahmen

    Read manual completely before beginning

    Lagerung

    -20 °C

    Informationen zur Lagerung

    Store kit components at -20°C unless specified otherwise. DO NOT USE past kit expiration date. Some vials contain a small amount of reagents. Spin tubes on pulse setting prior to opening.

    Haltbarkeit

    12 months
  • Target Alle Trastuzumab Produkte

    Trastuzumab

    Hintergrund

    Trastuzumab (Herceptin®) is indicated for the treatment of HER2-positive breast cancer, and adjuvant therapies for metastatic gastric cancer and gastroesophageal cancer. Serum concentration of Trastuzumab may predict some clinical outcome during therapy. It is also possible that the surveillance of circulating Trastuzumab concentration during therapy represents a direct factor for immunogenicity and some other side effects. Identification of biomarkers and risk factors for adverse drug reactions that might be related to serum concentrations, and maintaining the effective minimum concentration of Trastuzumab in order to potentially avoid some side effects, might be beneficial using a reliable method