Etanercept ELISA Kit
Kurzübersicht für Etanercept ELISA Kit (ABIN4886392)
Target
Alle Etanercept ProdukteReaktivität
Nachweismethode
Methodentyp
Detektionsbereich
Applikation
Proben
-
-
Untere Nachweisgrenze
- 1.56 ng/mL
-
Verwendungszweck
- Quantification of Etanercept in biological matrices
-
Analytische Methode
- Quantitative
-
Spezifität
- Etanercept (Enbrel)
-
Kreuzreaktivität (Details)
- sTNFa, TNFb, sTNFRI, sTNFRII prepared at 250 ng/mL were assayed and exhibited no crossreactivity or interference.
-
Sensitivität
- 1.5 ng/mL
-
Bestandteile
-
- Coated microtiter plate, 96 wells
- Calibrator diluent. - 1.8ml
- Calibrator 12ul 10X wash buffer - 25ml
- Assay buffer - 50ml
- 1000X detection reagent - 17ul TMB - 12ml
- TMB stop solution - 12ml
- Plate sealers - 3
-
Benötigtes Material
-
- Precision pipettes calibrated to deliver 5-1000μL
- Multi-channel pipette calibrated to deliver 50-200μL
- Plate shaker
- Disposable tips
- Vortex-Mixe
- Distilled or de-ionized water
- Microplate reader capable of reading 450nm with background subtraction
-
-
-
-
Applikationshinweise
- Optimal working dilution should be determined by the investigator.
-
Probenmenge
- 15 μL
-
Testdauer
- 2.5 h
-
Plattentyp
- Pre-coated
-
Protokoll
- The Etanercept ELISA kit is designed to measure free Etanercept with high specificity and sensitivity . This assay employs the sandwich enzyme immunoassay technique. A precoated anti-Etanercept 96 well plate is provided. Calibrator, quality control samples and test samples are pipetted into the appropriate wells. Etanercept present in biological matrices is bound by the immobilized capture antibody. After washing away any unbound substances, enzyme linked detection antibody is added to the wells. The plate is washed to remove any unbound antibody-enzyme reagent and a substrate solution is added to the wells for color development. The color development is proportional to the amount of Etanercept present in test samples and the concentration is calculated from the standard series.
-
Aufbereitung der Reagenzien
-
Prepare only the appropriate amount of required reagent on the day of use. Store all reagents as per instructions stated on the label. 1. Wash Buffer (1X) Preparation: Dilute wash buffer concentrate with ultra-pure water 1/10 before use (for example add 20 mL concentrate to 180 mL ultra-pure water). Mix well. 2. Detection Reagent (1X) Preparation: Dilute detection reagent with assay buffer 1/1000 before use (for example add 11 μL concentrate to 11 mL of assay buffer). Mix well. 3. Preparation of Calibrators: Prepare calibrators with concentrations ranging from 2500 ng/mL to 78 ng/mL. The following is an example calibrator curve.
-
Probennahme
- This kit is compatible with EDTA-plasma, heparinplasma and serum samples. Samples can be stored at or below -20 °C for up to 1 year.
-
Aufbereitung der Proben
-
Dilute calibrators and test samples 1/50 with assay buffer (for example add 5μL of prepared calibrator or sample to 245μL of assay buffer). Mix well. Do not store diluted samples.
-
Testdurchführung
-
This assay employs the sandwich enzyme immunoassay technique. Capture antibody is coated onto a 96 well microplate. Calibrator and test samples are pipetted into the appropriate wells. Etanercept present in biological matrices is bound by the immobilized anti-Etanercept antibody. After washing away any unbound substances, enzyme linked anti-Etanercept antibody is added to the wells. The plate is washed to remove any unbound antibody-enzyme reagent and a substrate solution is added to the wells for color development. The color development is proportional to the amount of Etanercept present in test samples. The color development is stopped and the intensity of the color is measured.
-
Ergebnisberechnung
-
- Construct a standard curve by plotting the absorbance obtained from each standard against concentration. Use a 4 or 5 parameter curve fit. Alternatively a log-log curve fit may be used. 2. The concentration of the unknowns can be read directly from this standard curve using the absorbance value for each sample. 3. Any sample undiluted or diluted still reading greater than the highest standard should be diluted appropriately with assay buffer and retested. If the samples have been diluted, the concentration determined from the standard curve must be multiplied by the dilution factor.
-
Testpräzision
-
Intra-assay precision: <10%
Inter-assay precision: <10% -
Beschränkungen
- Nur für Forschungszwecke einsetzbar
-
-
-
Konservierungsmittel
- Without preservative
-
Lagerung
- -20 °C
-
Informationen zur Lagerung
- -20C, 1 year
-
Haltbarkeit
- 12 months
-
-
- Etanercept
-
Hintergrund
- Etanercept (trade name Enbrel®) is a protein drug used to treat autoimmune diseases by adsorbing tumor necrosis factor (TNF, a soluble inflammatory cytokine). Etanercept is a fusion protein produced through expression of recombinant DNA by linking the extracellular ligand-binding portion of TNFRSF1B to the Fc component of human immunoglobulin G1 (IgG1). It reduces the effect of naturally present TNF, functioning as a decoy receptor that binds to TNF. Etanercept is indicated for the treatment of moderate to severe rheumatoid arthritis (RA), psoriatic arthritis, ankylosing spondylitis, and moderately to severely active polyarticular juvenile idiopathic arthritis. Serum concentration of Enbrel® may predict some clinical outcome during maintenance therapy.
-
Gen-ID
- 7124
Target Alle Etanercept Produkte
-