CRH ELISA Kit (Corticotropin Releasing Hormone)

Details for Product CRH ELISA Kit No. ABIN2851075, Anbieter: Anmelden zum Anzeigen
  • CRF
  • Gm1347
  • crf
  • crh
  • zgc:101864
  • CRH
  • corticoliberin
  • crhb
  • crh.1-a
  • crha
  • C1qrf
  • CTRP14
  • gliacolin
  • corticotropin releasing hormone
  • corticotropin releasing hormone b
  • corticotropin releasing hormone S homeolog
  • corticotropin releasing hormone L homeolog
  • complement component 1, q subcomponent-like 1
  • cryptothrix
  • Crh
  • CRH
  • crhb
  • crh
  • crh.S
  • crh.L
  • C1ql1
  • crhx
Kits mit alternativen Reaktivitäten:
Competition ELISA
15.625-1000 pg/mL
Untere Nachweisgrenze
15.625 pg/mL
Anmelden zum Anzeigen
Hersteller Produkt- Nr.
Anmelden zum Anzeigen
Verwendungszweck The kit is a competitive inhibition enzyme immunoassay technique for the in vitro quantitative measurement of CRH in mouse serum, plasma and other biological fluids.
Proben Plasma, Serum
Analytische Methode Quantitative
Nachweismethode Colorimetric
Spezifität This assay has high sensitivity and excellent specificity for detection of this index.
Kreuzreaktivität (Details) No significant cross-reactivity or interference between this index and analogues was observed. Note: Limited by current skills and knowledge, it is impossible for us to complete the cross- reactivity detection between this index and all the analogues, therefore, cross reaction may still exist.
Sensitivität 4.45 pg/mL
Produktmerkmale mouse Corticotropin Releasing Hormone (CRH) ELISA Kit
  • Pre-coated, ready to use 96-well strip plate
  • Standard (freeze dried)
  • Standard Diluent
  • Detection Reagent A
  • Detection Reagent B
  • Assay Diluent A
  • Assay Diluent B
  • TMB
  • Stop Solution
  • Wash Buffer (30X)
  • Plate sealer for 96 wells
  • Instruction manual
Benötigtes Material
  1. Microplate reader with 450 ± 10nm filter.
  2. Precision single or multi-channel pipettes and disposable tips.
  3. Eppendorf Tubes for diluting samples.
  4. Deionized or distilled water.
  5. Absorbent paper for blotting the microtiter plate.
  6. Container for Wash Solution.
Andere Bezeichnung CRH (CRH ELISA Kit Abstract)
Gen-ID 12918
UniProt Q8CIT0
Pathways Positive Regulation of Peptide Hormone Secretion, Hormone Activity, Negative Regulation of Hormone Secretion, cAMP Metabolic Process, Myometrial Relaxation and Contraction, Feeding Behaviour
Probenmenge 100 μL
Testdauer 1 - 4.5 h
Plattentyp Pre-coated

This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to the index has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled the index and unlabeled the index (Standards or samples) with the pre-coated antibody specific to the index. After incubation the unbound conjugate is washed off. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of the index in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of the index in the sample.

  • Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
  • Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
  • CV(%) = SD/meanX100
  • Intra-assay: CV<10%
  • Inter-assay: CV<12%
Beschränkungen Nur für Forschungszwecke einsetzbar
Vorsichtsmaßnahmen The Stop Solution suggested for use with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
Handhabung The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Lagerung 4 °C,-20 °C
Informationen zur Lagerung The Assay Plate, Standard, Detection Reagent A and Detection Reagent B should be stored at -20°C upon being received. After receiving the kit , Substrate should be always stored at 4°C.Other reagents are kept according to the labels on vials. But for long term storage, please keep the whole kit at -20°C. The unused strips should be kept in a sealed bag with the desiccant provided to minimize exposure to damp air. The test kit may be used throughout the expiration date of the kit (six months from the date of manufacture). Opened test kits will remain stable until the expiring date shown, provided it is stored as prescribed above.
Haltbarkeit 12 months
Haben Sie etwas anderes gesucht?