UCN3 Reaktivität: Human Colorimetric Sandwich ELISA 31.25 pg/mL - 2000 pg/mL Plasma, Serum
Produktnummer ABIN5519133
  • Target Alle UCN3 ELISA Kits anzeigen
    UCN3 (Urocortin 3 (UCN3))
    Reaktivität
    • 4
    • 2
    • 2
    • 1
    Human
    Nachweismethode
    Colorimetric
    Methodentyp
    Sandwich ELISA
    Detektionsbereich
    31.25 pg/mL - 2000 pg/mL
    Untere Nachweisgrenze
    31.25 pg/mL
    Applikation
    ELISA
    Verwendungszweck
    The kit is a sandwich enzyme immunoassay technique for the in vitro quantitative measurement in various sample types.
    Proben
    Plasma, Serum
    Analytische Methode
    Quantitative
    Sensitivität
    18.75 pg/mL
    Bestandteile
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Reference Standard & Sample Diluent
    • Biotinylated Detection Antibody (100 x concentrate)
    • HRP Conjugate (100 x concentrate)
    • Biotinylated Detection Antibody Diluent
    • HRP Conjugate Diluent
    • Substrate Reagent
    • Stop Solution
    • Wash Buffer (25 x concentrate)
    • Instruction manual
    Featured
    Zu unserem meistverkauften UCN3 ELISA Kit
    Top Product
    Discover our top product UCN3 ELISA Kit
  • Probenmenge
    100 µL
    Testdauer
    1 - 4.5 h
    Plattentyp
    Pre-coated
    Protokoll
    1. Add 100 µL standard or sample to each well. Incubate for 90 min at 37 °C.
    2. Remove the liquid. Add 100 µL Biotinylated Detection Antibody. Incubate for 1 hour at 37 °C.
    3. Aspirate and wash 3 times.
    4. Add 100 µL HRP Conjugate. Incubate for 30 min at 37 °C.
    5. Aspirate and wash 5 times.
    6. Add 90 µL Substrate Reagent. Incubate for 15 min at 37 °C.
    7. Add 50 µL Stop Solution. Read at 450 nm immediately.
    8. Calculation of results.
    Aufbereitung der Reagenzien
    1. Bring all reagents to room temperature (18~25 °C) before use. Follow the Microplate reader manual for set-up and preheat it for 15 min before OD measurement.
    2. Wash Buffer: Dilute 30 mL of Concentrated Wash Buffer with 720 mL of deionized or distilled water to prepare 750 mL of Wash Buffer.Note: if crystals have formed in the concentrate, warm it in a 40 °C water bath and mix it gently until the crystals have completely dissolved
    3. Standard working solution: Centrifuge the standard at 10,000xg for 1 min. Add 1.0 mL of Reference Standard &Sample Diluent, let it stand for 10 min and invert it gently several times. After it dissolves fully, mix it thoroughly with a pipette. This reconstitution produces a working solution of 2000 pg/mL. Then make serial dilutions as needed. The recommended dilution gradient is as follows: 2000, 1000, 500, 250, 125, 62.5, 31.25, 0 pg/mL. Dilution method: Take 7 EP tubes, add 500 μLof Reference Standard & Sample Diluent to each tube. Pipette 500 μLof the 2000 pg/mL working solution to the first tube and mix up to produce a 1000 pg/mL working solution. Pipette 500 μLof the solution from the former tube into the latter one according to these steps. The illustration below is for reference. Note: the last tube is regarded as a blank. Don't pipette solution into it from the former tube.
    4. Biotinylated Detection Antibody working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Centrifuge the stock tube before use, dilute the 100x Concentrated Biotinylated Detection Antibody to 1xworking solution with Biotinylated Detection Antibody Diluent.
    5. Concentrated HRP Conjugate working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Dilute the 100x Concentrated HRP Conjugate to 1x working solution with Concentrated HRP Conjugate Diluent.
    Beschränkungen
    Nur für Forschungszwecke einsetzbar
  • Lagerung
    4 °C/-20 °C
    Informationen zur Lagerung
    The unopened kit can be stored at 4℃ for 1 month. If the kit is not used within 1 month, store the items separately according to the conditions since the kit is received.
  • Target Alle UCN3 ELISA Kits anzeigen
    UCN3 (Urocortin 3 (UCN3))
    Andere Bezeichnung
    UCN3(Urocortin 3) (UCN3 Produkte)
    Synonyme
    SCP ELISA Kit, SPC ELISA Kit, UCNIII ELISA Kit, ucn3 ELISA Kit, RGD1562213 ELISA Kit, urocortin 3 ELISA Kit, urocortin 3 S homeolog ELISA Kit, UCN3 ELISA Kit, Ucn3 ELISA Kit, ucn3 ELISA Kit, ucn3.S ELISA Kit
    Pathways
    Positive Regulation of Peptide Hormone Secretion
Sie sind hier:
Kundenservice