Telefon:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@antikoerper-online.de

PLA2G2A ELISA Kit

PLA2G2A Reaktivität: Human Colorimetric Sandwich ELISA 78-5000 pg/mL Cell Culture Supernatant, Plasma, Serum
Produktnummer ABIN456378
  • Target Alle PLA2G2A ELISA Kits anzeigen
    PLA2G2A (Phospholipase A2, Group IIA (Platelets, Synovial Fluid) (PLA2G2A))
    Reaktivität
    • 3
    • 2
    • 1
    • 1
    Human
    Nachweismethode
    Colorimetric
    Methodentyp
    Sandwich ELISA
    Detektionsbereich
    78-5000 pg/mL
    Untere Nachweisgrenze
    78 pg/mL
    Applikation
    ELISA
    Verwendungszweck
    This immunoassay kit allows for the specific measurement of human secreted phospholipase A2, sPLA2 concentrations in cell culture supernates, serum and plasma.
    Proben
    Cell Culture Supernatant, Serum, Plasma
    Analytische Methode
    Quantitative
    Spezifität
    This assay recognizes recombinant and natural human sPLA2.
    Kreuzreaktivität (Details)
    No significant cross-reactivity or interference was observed.
    Sensitivität
    < 7.8 pg/mL
    The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest detectable concentration that could be differentiated from zero.
    Produktmerkmale
    Homo sapiens,Human,Phospholipase A2, membrane associated,GIIC sPLA2,Group IIA phospholipase A2,Non-pancreatic secretory phospholipase A2,NPS-PLA2,Phosphatidylcholine 2-acylhydrolase 2A,PLA2G2A,
    Bestandteile
    Reagent (Quantity): Assay plate (1), Standard (2), Sample Diluent (1x20ml), Assay Diluent A (1x10ml), Assay Diluent B (1x10ml), Detection Reagent A (1x120µl), Detection Reagent B (1x120µl), Wash Buffer(25 x concentrate) (1x30ml), Substrate (1x10ml), Stop Solution (1x10ml)
    Featured
    Zu unserem meistverkauften PLA2G2A ELISA Kit
    Top Product
    Discover our top product PLA2G2A ELISA Kit
  • Probenmenge
    100 μL
    Plattentyp
    Pre-coated
    Protokoll
    This assay employs the quantitative sandwich enzyme immunoassay technique. A antibody specific for sPLA2 has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and any sPLA2 present is bound by the immobilized antibody. An enzyme-linked antibody specific for sPLA2 is added to the wells. Following a wash to remove any unbound antibody-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of sPLA2 bound in the initial step. The color development is stopped and the intensity of the color is measured.
    Aufbereitung der Reagenzien

    Bring all reagents to room temperature before use. Wash Buffer - If crystals have formed in the concentrate, warm to room temperature and mix gently until the crystals have completely dissolved. Dilute 20 mL of Wash Buffer Concentrate into deionized or distilled water to prepare 500 ml of Wash Buffer. Standard - Reconstitute the Standard with 1.0 mL of Sample Diluent. This reconstitution produces a stock solution of 2,000 pg/mL. Allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making serial dilutions. The undiluted standard serves as the high standard (2,000 pg/mL). The Sample Diluent serves as the zero standard (0 pg/mL). Detection Reagent A and B - Dilute to the working concentration specified on the vial label using Assay Diluent A and B (1:100), respectively.

    Probennahme
    Cell culture supernates - Remove particulates by centrifugation and assay immediately or aliquot and store samples at ≤ -20 °C. Avoid repeated freeze-thaw cycles. Serum - Use a serum separator tube (SST) and allow samples to clot for 30 minutes before centrifugation for 15 minutes at approximately 1000 x g. Remove serum and assay immediately or aliquot and store samples at -20 °C. Plasma - Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 15 minutes at 1000 x g at 2 - 8 °C within 30 minutes of collection. Store samples at ≤ -20 °C. Avoid repeated freeze-thaw cycles. Note: Citrate plasma has not been validated for use in this assay.
    Testdurchführung

    Allow all reagents to reach room temperature. Arrange and label required number of strips.
    1. Prepare all reagents, working standards and samples as directed in the previous sections.
    2. Add 100 uL of Standard, Control, or sample per well. Cover with the adhesive strip. Incubate for 2 hours at 37 °C.
    3. Remove the liquid of each well, don’t wash.
    4. Add 100 uL of Detection Reagent A to each well. Incubate for 1 hour at 37°C. Detection Reagent A may appear cloudy. Warm to room temperature and mix gently until solution appears uniform. 3
    5. Aspirate each well and wash, repeating the process three times for a total of three washes. Wash by filling each well with Wash Buffer (350 uL) using a squirt bottle, multi-channel pipette, manifold dispenser or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot it against clean paper towels.
    6. Add 100 uL of Detection Reagent B to each well. Cover with a new adhesive strip.Incubate for 1 hours at 37 °C.
    7. Repeat the aspiration/wash as in step
    5. 8. Add 90 uL of Substrate Solution to each well. Incubate for 30 minutes at room temperature. Protect from light.
    9. Add 50 uL of Stop Solution to each well. If color change does not appear uniform, gently tap the plate to ensure thorough mixing.
    10. Determine the optical density of each well within 30 minutes, using a microplate reader set to 450 nm.
    Important Note:
    1. The wash procedure is critical. Insufficient washing will result in poor precision and falsely elevated absorbance readings.
    2. It is recommended that no more than 32 wells be used for each assay run if manual pipetting is used since pipetting of all standards, specimens and controls should be completed within 5 minutes. A full plate of 96 wells may be used if automated pipetting is available.
    3. Duplication of all standards and specimens, although not required, is recommended.
    4. When mixing or reconstituting protein solutions, always avoid foaming.
    5. To avoid cross-contamination, change pipette tips between additions of each standard level, between sample additions, and between reagent additions. Also, use separate reservoirs for each reagent.
    6. To ensure accurate results, proper adhesion of plate sealers during incubation steps is necessary. 4

    Ergebnisberechnung

    Average the duplicate readings for each standard, control, and sample and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a standard curve by plotting the mean absorbance for each standard on the y-axis against the concentration on the x-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the sPLA2 concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.

    Beschränkungen
    Nur für Forschungszwecke einsetzbar
  • Handhabung
    1. The kit should not be used beyond the expiration date on the kit label.
    2. Do not mix or substitute reagents with those from other lots or sources.
    3. If samples generate values higher than the highest standard, further dilute the samples with the Assay Diluent and repeat the assay. Any variation in standard diluent, operator, pipetting technique, washing technique,incubation time or temperature, and kit age can cause variation in binding.
    4. This assay is designed to eliminate interference by soluble receptors, ligands, binding proteins, and other factors present in biological samples. Until all factors have been tested in the Immunoassay, the possibility of interference cannot be excluded.
    Lagerung
    4 °C/-20 °C
    Informationen zur Lagerung
    The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20 °C upon being received. The other reagents can be stored at 4 °C.
  • Target Alle PLA2G2A ELISA Kits anzeigen
    PLA2G2A (Phospholipase A2, Group IIA (Platelets, Synovial Fluid) (PLA2G2A))
    Andere Bezeichnung
    PLA2G2A (PLA2G2A Produkte)
    Synonyme
    MOM1 ELISA Kit, PLA2 ELISA Kit, PLA2B ELISA Kit, PLA2L ELISA Kit, PLA2S ELISA Kit, PLAS1 ELISA Kit, sPLA2 ELISA Kit, EF ELISA Kit, Mom1 ELISA Kit, Pla2 ELISA Kit, sPla2-IIA ELISA Kit, phospholipase A2 group IIA ELISA Kit, phospholipase A2, group IIA (platelets, synovial fluid) ELISA Kit, PLA2G2A ELISA Kit, Pla2g2a ELISA Kit
    Hintergrund
    Phospholipase A2 (PLA2) catalyzes the hydrolysis of phospholipids at the sn-2 position yielding a free fatty acid and a lysophospholipid. The release of arachidonic acid from membrane phospholipids by PLA2 is believed to be a key step in the control of eicosanoid production within the cell.
    Pathways
    Stem Cell Maintenance, Inositol Metabolic Process
Sie sind hier:
Kundenservice