Telefon:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@antikoerper-online.de

Total Polyphenols Quantification Assay Kit

Q Colorimetric Beverages, Food Quantitative Uncoated
Produktnummer ABIN7384588
  • Nachweismethode
    Colorimetric
    Detektionsbereich
    3.5-25 μg/mL
    Untere Nachweisgrenze
    3.5 μg/mL
    Applikation
    Quantification (Q)
    Verwendungszweck
    The kit is designed for the rapid determination of polyphenols in various samples.
    Proben
    Beverages, Food
    Analytische Methode
    Quantitative
    Spezifität
    Specific for phenol content determination.
    Produktmerkmale
    Phenolic Quantification Assay is based on Folin-Ciocalteu method. The FC reagent contains phosphomolybdic/ phosphotungstic acid complexes1. The method relies on the transfer of electrons in alkaline medium from phenolic compounds to form a blue chromophor
    Bestandteile
    Plate and the reagents neccesary to perform the assay.
    Benötigtes Material
    Pipettes, reaction tubes, plate reader.
  • Kommentare

    1 ml (microassay)/ 20μL (microplate)

    Testdauer
    15 min
    Plattentyp
    Uncoated
    Aufbereitung der Reagenzien

    Reagent A -FC reagent-needs to be diluted (1:10)

    Testdurchführung

    MICROASSAY: 1.Pipette 1 mL of each standard, unknown sample or diluted sample replicate into separate clean test tubes. Refer to the Table 1 (see booklet) as a guide for diluting the standard phenol. For the diluent, use the same buffer as in the samples. 2.Add 1 mL of Reagent A -Folin-Ciocalteu Reagent previously diluted (1:10) in deionized water- to each tube. 3.Add 1 mL of Reagent B (Alkaline Working Solution) to each tube and measure the absorbance at 700 nm immediately. 4.Measure the absorbance of these standards, blanks and unknown samples at 700 nm. MICROPLATE: 1. Prepare standards containing a range of 1 to 10 μg phenol (Gallic acid) to a volume of 200 μL. Pipette 20 μL of each standard and unknown sample or diluted sample replicate into a microplate well. Refer to the Table 2 (see booklet) as a guide for diluting the standard phenol. For the diluent, use the same buffer as in the samples. 2.To each well, add 100 μL of Reagent A -FC reagent-previously diluted (1:10) in deionized water. 3.Add 80 μL ofReagent B (Alkaline Working Solution) to each well and measure the absorbance at 700 nm on a plate reader immediately. 4.Measure the absorbance of these standards, blanks and unknown samples at 700 nm.

    Ergebnisberechnung

    1.If the spectrophotometer or microplate reader was not zeroed with the blank, then substract the average blank value from the standard and unknown sample values. 2.Create a standard curve by plotting A700nm (y-axis) vs standard, μg (x-axis). Determine the unknown sample concentration using the standard curve. 3.Standard curve example for microplate assay procedure is shown in Figure 2 (see images). 4.Dilute the unknown samples until they reach an absorbance within the limits of the standard curve.

    Beschränkungen
    Nur für Forschungszwecke einsetzbar
  • Konservierungsmittel
    Sodium azide
    Vorsichtsmaßnahmen
    This product contains Sodium azide: a POISONOUS AND HAZARDOUS SUBSTANCE which should be handled by trained staff only.
    Lagerung
    RT/4 °C
    Informationen zur Lagerung
    FC Reagent A: RT, FC Reagent B: RT, FC Standard: 4 °C