Telefon:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@antikoerper-online.de

Motilin ELISA Kit

MLN Reaktivität: Human Colorimetric Competition ELISA 15.625 pg/mL - 1000 pg/mL Plasma, Serum
Produktnummer ABIN1029312
  • Target Alle Motilin (MLN) ELISA Kits anzeigen
    Motilin (MLN)
    Reaktivität
    • 7
    • 3
    • 2
    • 2
    • 2
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    Human
    Nachweismethode
    Colorimetric
    Methodentyp
    Competition ELISA
    Detektionsbereich
    15.625 pg/mL - 1000 pg/mL
    Untere Nachweisgrenze
    15.625 pg/mL
    Applikation
    ELISA
    Verwendungszweck
    The kit is a competitive inhibition enzyme immunoassay technique for the in vitro quantitative measurement of MTL in human serum, plasma and other biological fluids.
    Proben
    Plasma, Serum
    Analytische Methode
    Quantitative
    Spezifität
    This assay has high sensitivity and excellent specificity for detection of this index.
    Kreuzreaktivität (Details)
    No significant cross-reactivity or interference between this index and analogues was observed. Note: Limited by current skills and knowledge, it is impossible for us to complete the cross- reactivity detection between this index and all the analogues, therefore, cross reaction may still exist.
    Sensitivität
    4.28 pg/mL
    Bestandteile
    • Pre-coated, ready to use 96-well strip plate
    • Standard (freeze dried)
    • Standard Diluent
    • Detection Reagent A
    • Detection Reagent B
    • Assay Diluent A
    • Assay Diluent B
    • TMB
    • Stop Solution
    • Wash Buffer (30X)
    • Plate sealer for 96 wells
    • Instruction manual
    Benötigtes Material
    1. Microplate reader with 450 ± 10nm filter.
    2. Precision single or multi-channel pipettes and disposable tips.
    3. Eppendorf Tubes for diluting samples.
    4. Deionized or distilled water.
    5. Absorbent paper for blotting the microtiter plate.
    6. Container for Wash Solution.
    Featured
    Zu unserem meistverkauften MLN ELISA Kit
    Top Product
    Discover our top product MLN ELISA Kit
  • Probenmenge
    50 μL
    Testdauer
    1 - 4.5 h
    Plattentyp
    Pre-coated
    Protokoll
    1. Prepare all reagents, samples and standards
    2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C
    3. Aspirate and wash 3 times
    4. Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
    5. Aspirate and wash 5 times
    6. Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
    7. Add 50µL Stop Solution. Read at 450 nm immediately.
    Testdurchführung

    This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to the index has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled the index and unlabeled the index (Standards or samples) with the pre-coated antibody specific to the index. After incubation the unbound conjugate is washed off. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of the index in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of the index in the sample.

    Testpräzision
    • Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
    • Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
    • CV(%) = SD/meanX100
    • Intra-assay: CV<10%
    • Inter-assay: CV<12%
    Beschränkungen
    Nur für Forschungszwecke einsetzbar
  • Vorsichtsmaßnahmen
    The Stop Solution suggested for use with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
    Handhabung
    The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
    Lagerung
    4 °C,-20 °C
    Informationen zur Lagerung
    The Assay Plate, Standard, Detection Reagent A and Detection Reagent B should be stored at -20°C upon being received. After receiving the kit , Substrate should be always stored at 4°C.Other reagents are kept according to the labels on vials. But for long term storage, please keep the whole kit at -20°C. The unused strips should be kept in a sealed bag with the desiccant provided to minimize exposure to damp air. The test kit may be used throughout the expiration date of the kit (six months from the date of manufacture). Opened test kits will remain stable until the expiring date shown, provided it is stored as prescribed above.
    Haltbarkeit
    12 months
  • Target Alle Motilin (MLN) ELISA Kits anzeigen
    Motilin (MLN)
    Andere Bezeichnung
    Motilin (MTL) (MLN Produkte)
    Synonyme
    MLN ELISA Kit, motilin ELISA Kit, MLN ELISA Kit, Mln ELISA Kit
    Hintergrund
    Alternative name: MLN, MAP, Promotilin, Motilin-associated peptide
    Gen-ID
    4295
    UniProt
    P12872
    Pathways
    Hormone Activity
Sie sind hier:
Kundenservice